赛默飞世尔科技为您服务
联系销售
联系技术与支持
点击放大
HP-360 GM探头
The HP-360 hand probe provides a sensitive beta detector, featuring a "Pancake" GM tube with a thin mica window. It is designed for contamination surveys on personnel, table tops, floors, equipment, etc. 
 This detector permits useful beta sensitivities down to 40 keV and is alpha sensitive above 3 MeV. The background sensitivity to 137Cs is approximately 3600 cpm/mR/hr and it is capable of seeing low energy photons down to approximately 10 keV. 

This detector is also available in a "Smart" configuration (Model SHP-360) for use with Thermo Scientific's new generation of smart instruments like the Model E-600.

Application : Alpha/Beta/Gamma surveys; frisking
Detector Type : GM, non compensated
Operating Voltage : 900 V +/- 50V
Dead Time : 50 µs nominal
Mica Window Size : 1.75" diameter (4.4 cm)
Mica Window Thickness : 1.4 to 2.0 mg/cm^2
Background Sensitivity : ~ 3,600 cpm / mR/h (Cs137)
Beta/Gamma Efficiency : ~ 22% Cs137; ~ 16% Co60
Beta Efficiency (4Pi) : ~ 32% Sr90/Y90; ~ 15% Tc99; ~ 6% C14
Alpha Efficiency (4Pi) : ~ 25% Am241
Operating Temp : -22° to +140°F (-30° to +60° C)
Housing : ABS Plastic
Connector : BNC
Size : 9.75" long x 2.68" wide x 2.75" high (24.8 x 6.8 x 7.0 cm)
Weight : 0.5 lbs (0.2 kg)