Pascal 240孔径分析仪
This Pascal 240 measures  pores down to the lower mesopore region and operates in the classical range from atmospheric pressure up to 200 MPa.  Developed to meet the sophisticated analytical requirements of laboratories working with modern materials.
The Pascal 240 porosimeter operates in the classical range from atmospheric pressure up to 200 MPa. It takes over where the Pascal 140 leaves off to measure pores down to the lower mesopore region. This unit is versatile, high-pressure porosimeter working up to 200 MPa. Its large autoclave can accept dilatometers of different sizes, enabling the user to characterize samples of different porous structures (i.e. homogeneous, heterogeneous, low or high porosity). The accuracy and reproducibility are the utmost thanks to the PASCAL system. It requires a sample preparation unit to degas the sample before the analysis and to fill the dilatometer with fresh mercury. The dilatometer preparation unit can be the Pascal 140 or a Duplex Dilatometer filling device.

Speed and resolution to match any requirement

The Pascal 240 incorporates a totally new pressurization system, developed to meet the sophisticated analytical requirements of laboratories working with modern materials. The pressurization system uses a reversible pump operating continuously and a new type of pressure multiplier. This permits a perfect control of the pressurization speed during the analysis. The key features of this system are the extremely high maximum speed of pressurization and the immediate acceleration or deceleration response according to the real mercury penetration speed in the sample porous matrix. At the maximum speed, the instrument can provide a penetration curve up to 200 MPa in about 9 minutes, permitting an incredible productivity in terms of analyses per hour, as required in quality control labs. The easy-to-use control panel with a liquid crystal display allows operating the unit even without a PC connected.

Main features

  • Automatic autoclave opening/closing           
  • Autoclave oil filling by pumping system           
  • Up to 2500 experimental points in every run           
  • All dilatometers are accepted           
  • Only 9 minutes to reach the maximum pressure
  • IMPORTANT NOTE:  The PASCAL 240 not currently available in US and Canada

     

    资源文章
    更多>>
    Pressure range :0.1 - 200 MPa
    Pressure resolution :0.01 MPA (up to 100 MPa); 0.1 MPA (100 - 200 MPa)
    Pressure accuracy :better than 0.2%
    Maximum detectable volume :500 mm3 (resolution: 0.1 mm3); 2000 mm3 (resolution: 0.5 mm3)
    Measuring ranges: pore radius :7500 - 3.7 nm
    Measuring ranges: particle diameter :40 - 0.015 micrometers