| Measuring Ranges : |
Full Scale Ranges from 0-5 ppm S (w/w) to 0-9000 ppm S (w/w) |
| Repeatability : |
For full scale ranges > or equal to 15 ppm S ±1% of full scale (signal average times may vary with full scale measuring range); For full scale ranges <15 ppm S ±2% of full scale (signal average times may vary with full scale measuring range) |
| Linearity : |
±1% of full scale |
| Response Time : |
Semi-continuous, outputs updated every 10 seconds, 3-5 minutes to 90% of new value |
| Number of Process Streams : |
Dual streams with auto stream select (optional) |
| Calibration : |
Automatic or manual |
| Analog Outputs : |
4-20 mA DC for each stream (optional) |
| Alarm Outputs : |
One global dry contact triggered by one or more of the following; Low sample flow alarm (optional); low detector flow alarm; Oven/Pyrolyzer temperature fault; Injection value fault; Purge failure; Calibration fault; Detector temperature fault; Detector lamp voltage fault; One out of service dry contact triggered by; Analyzer in calibration; Suspension of analyzer |
| Analog Inputs : |
Optional 4-20 mA DC inputs from density meter for automatic density compensation of ppm S (w/w); Optional 4-20 mA DC input from sample flowmeter |
| Digital Data Communications : |
Dual channel with the following optional configurations; RS-232 Modbus & RS-485 Modbus; Dual channel RS-485 Modbus; TCP/IP encapsulated Modbus & RS-485 Modbus |
| Local MMI : |
Status of all analyzers parameters (e.g., furnace & oven temperatures, PMT and lamp voltage, detector flow rate, etc.) and analytical results available on front mounted displays, push button menu access, hazardous area classification remains intact while operating local display |
| SOLA II Modbus Remote Interface : |
Complete remote control of SOLA; automatic logging of analysis results and analyzer parameters; Communication to SOLA II via serial or TCP/IP encapsulated Modbus enables remote diagnostics via modem |
| SOLAWeb Remote Interface : |
Complete remote control of SOLA; Ability to download 24 hours of analysis results and analyzer parameters; Communication to SOLA II via local area network (TCP/IP) enables remote diagnostics via modem |
| Ambient Temperature : |
+12°C to +40°C (+54°F to +104°F) |
| Power : |
110 VAC, 50/60 Hz at 2000 watts; 220 VAC, 50/60 Hz at 2000 watts |
| Instrument Air : |
80 psig (5.5 barg), 8 SCFM, Oil Free, -40°C (-40°F) dew point |
| Zero Grade Air : |
80 psig (5.5 barg), 200-300 SCCM |
| Zone 1; Div 1 Congigurations : |
1581.15 mm (62.25 in) high x 647.70 mm (25.50 in) wide x 476.25 mm (18.75 in) deep |
| Zone 2; Div. 2 Configurations : |
1104.39 mm (43.48 in) high x 647.70 mm (25.50 in) wide x 476.25 mm (18.75 in) deep |
| Certification : |
(Built to) NEC Class 1 Division 2, Group B, C, D; (Built to) NEC Class 1 Division 1 (optional) Group B, C, D; CSA with associated "C/US Mark" Class 1, Division 2, Group B, C, D (pending); CSA with associated "C/US Mark" Class 1, Division 1 (optional), Group B, C, D (pending), ATEX Zone 2 EEx p IIC T2 (T3, T4 optional); ATEX Zone 1 (optional), EEx p IIC T2 (T3, T4 optional), CE Mark |